The CRO
measuring instrument may sound very familiar, as it is a very useful device. It
is used for measurement of voltages (AC/DC) and display of waveforms by
providing information on time duration, frequency, and their shapes.
Features of
CRO
Below are
listed the various features of the CRO:
• It allows
voltage (AC/DC) amplitude measurement and time period measurement from the
waveform displayed on screen.
• Dual trace
CRO allows the user to see two traces at a time on two different channels for
comparison.
• Two sets
of controls provide the facility to show time period differences, amplitude
differences, and shape/distortion comparison.
• Storage
oscilloscopes allow storage of waveforms for later analysis.
• Storage
facility is very useful since it provides a cursor function, which shows the
value of a measured variable at a particular instance.
Operating
CRO
To operate
the CRO, perform the following procedure:
• The power
on switch is provided for on/off control.
• The
measurement probe provided consists of two leads – one connected to the signal
and the other, ground probe, connected to the ground of circuit.
• Turn on
the CRO.
• Check the
integrity of leads and the CRO by connecting the I/P probe to a test socket of
5 V square wave signal.
• While
checking non-isolated signals (that are earthed) do not connect ground/earth to
the CRO, else it may create short-circuit at the input signal.
• Adjust
both channels’ vertical axis by placing AC\DC\GND signal in GND position.
• Place the
function switch in the suitable signal function as required (AC\DC).
• Check the
test probe selection, i.e., divide by 1 or 10 that allows signal attenuation.
• The
intensity knob is used to vary the brightness of the trace.
• The focus
knob is used to change the sharpness of the trace displayed.
• The Y
shift allows you to shift the waveform displayed vertically (up/down).
• The X
shift allows you to shift the waveform displayed horizontally left or right.
• The
Volts/Div switch is used to vary the magnitude of the voltage variable displayed
on the screen. It is calibrated in Volts/Div of the vertical scale.
A control
knob is provided in the center to adjust amplitude between calibrated settings.
• To find
the amplitude of a signal multiply the Y-axis reading with the Volts/Div setting.
• The
Time/Div is used to control the span of the X-axis.
• Physical
markings between two points can be used to calculate the time span.
The same
time span can be used to measure the frequency of the waveform displayed.
• A control
knob is provided at the center for the same purpose as in Volts/Div.
• To find
the time duration of a waveform, measure the signal span reading difference.
When this is multiplied by Times/Div it will give the time duration of the
signal.
No comments:
Post a Comment